Timing Basics
|
-there are 3 peaks on a balloon pump waveform
-#1, PSP, pts peak systolic BP -#3, APSP, assisted pts systolic BP, which reflects unloading -#2, is the highest peak, it follows the first peak or the PSP, it is the balloon doing is job -the first goal of proper timing is to make sure that the assisted systole is lower than the pts own systole |
|
Timing Basics cont
|
-after the dicrotic notch, which is the point at which the balloon inflates, the waveform shoots up
-this is the pressure generated in the aorta as the balloon inflates -the inflation is happening during diastole, this high part of the waveform is called the augmented diastolic pressure ADP or the #2 peak |
|
Balloon Aortic End Diastolic Pressure
|
-the lowest part of the waveform BAEDP
-this point should always be LOWER than the pts own diastolic pressure -this is the second goal of proper timing, to lower diastolic resistance in the arteries -with a machine set at a ration of 1:2, the assisted peak systolic BP should be lower than the pts systole -the BAEDP should be lower than the patients diastole |
|
Timing problems
|
-two bad timing problems are EARLY INFLATION and LATE DEFLATION
|
|
Early inflation
|
-the inflation knob id turned too far to the left, and the inflation wave comes before the dicrotic notch
-this means that the balloon is inflating before the aortic valve closes, pumping backwards into the left ventricle -the inflation point (dicrotic notch) is way out ahead of where it ought to be |
|
Late deflation
|
-is when the balloon remains inflated too long, the heart is trying to pump against an inflated balloon
-the deflation knob is turned too far to the right, it should be moved back towards the center -make sure the BAEDP is lower than the pts diastolic pressure |