4 things that get examined for radiopharmacy quality control
|
radionuclide purity
radiochemical purity chemical prutiy sterility and free of pyogens |
|
what are you looking for when testing radionuclide purity - how is test done
|
Mo breakthrough - test with lead shield for quanities >0.15 uCi
Alumina (Al2O3) breakthrough - test with indicator paper, which turns red (may result in clumping with liver and lung uptake), shouldu be <10ug/mL of Tc99 eluate |
|
how is particle size tested
|
hemocytometer
|
|
preferred size of MAA when doing VQ scan
|
150 microns (90% should be btwn 10-90 microns)
|
|
how is steritlity tested
|
inject 3 rabbits intravenously and test for gram-negative endotoxiins
limulus amebocyte lysate test |
|
what QC entity is tested daily
|
constancy
|
|
what QC entitiy is checked annually
|
accuracy
|
|
what QC entitiy is checked at installation
|
geometry
|
|
what QC entitiy is checked at installation and quarterly
|
linearity
|
|
how is radiochemical purity tested
|
with TLC
|
|
abnormal linearity test findings
|
barrel and pincushion deformities
|