- Shuffle
Toggle OnToggle Off
- Alphabetize
Toggle OnToggle Off
- Front First
Toggle OnToggle Off
- Both Sides
Toggle OnToggle Off
Front
How to study your flashcards.
Right/Left arrow keys: Navigate between flashcards.right arrow keyleft arrow key
Up/Down arrow keys: Flip the card between the front and back.down keyup key
H key: Show hint (3rd side).h key
![]()
PLAY BUTTON
![]()
PLAY BUTTON
![]()
57 Cards in this Set
- Front
- Back
|
Which way are detectors situated on the table
|
small elements along the z-axis (patient table) direction.
|
|
Are detectors set up so they are seperate elements in the X-Y plane
|
yes
|
|
How long does it take to scan a 40cm volume length
|
30 seconds
|
|
What is required to create a good quality 3D resolution image
|
near isotropic resolution
|
|
What is a CT image composed of
|
Figure 1. Sample CT image. A CT image is com-
posed of pixels (picture elements). Each pixel on the image represents the average x-ray attenuation in a small volume (voxel) that extends through the tissue section. (In this example, the pixel size is exaggerated. In addition, in a real CT image, all tissues within a single pixel would be the same shade of gray.) w width |
|
Why is the CT constrast so much better than plain film
|
because of the elimination of scatter
|
|
What is the fundamental way a CT works
|
Fundamentally, a CT scanner makes many mea-
surements of attenuation through the plane of a finite-thickness cross section of the body. The system uses these data to reconstruct a digital im- age of the cross section, with each pixel in the im- age representing a measurement of the mean at- tenuation of a boxlike element (a voxel) that ex- tends through the thickness of the section |
|
What is an attenuation measurement in CT
|
An attenuation measurement quantifies the
fraction of radiation removed in passing through a given amount of a specific material of thickness |
|
What is being measured by the detector
|
straight-line path through the patient from
the tube focal spot to a detector is the sum of the individual attenuations of all materials along the path. If it is assumed that the ray path through the tissue is broken up into incremental voxel thick- nesses |
|
How does the CT know how much attenuation to give each voxel in a straight path
|
The image reconstruction process derives the
average attenuation coefficient () values for each voxel in the cross section by using many rays from many different rotational angles around the cross section |
|
What is the basic fundamental cause of attenuation within each voxel
|
voxel attenuation increases with the density and the atomic num-
bers of tissues averaged through the volume of the voxel and declines with increasing x-ray energy. |
|
What is done to the date used to make up the density of the voxel
|
Be-
cause this process generates a blurred image, the data from each projection are mathematically al- tered (filtered) prior to back-projection, eliminat- ing the intrinsic blurring effect. |
|
How does the computer label the density of a voxel
|
by numbers
|
|
What does a second generation scanner look like
|
, one
early design used three detectors each displaced by 1°. Since each detector viewed the x-ray tube at a different angle, a single translation produced three projections. Hence, the system could rotate 3° to the next projection rather than 1° and had to make only 60 translations instead of 180 to ac- quire a complete section (Fig 5). |
|
What did a 1st generation scanner look like
|
A single projection
was acquired by moving the tube and detector in a straight-line motion (translation) on opposite sides of the patient (Fig 3). To acquire the next projection, the frame rotated 1°, then translated in the other direction. This process of translation and rotation was repeated until 180 projections were obtained |
|
What is the gantry
|
The gantry is a moveable frame that contains the x-ray tube including
collimators and filters, detectors, data acquisition system (DAS), rotational components including slip ring systems and all associated electronics such as gantry angulation motors and positioning laser lights. |
|
What does slip ring technology allow the CT to do
|
Slip ring technology eliminated the
need for cables and allows continuous rotation of the gantry components. The inclusion of slip ring technology into a CT system allows for continuous scanning without interference of cables for power |
|
What is the gantry aperture
|
the opening through which the patient passes
|
|
How does the scanning process work with newer scanners
|
The development of spiral/helical CT allows
continuous scanning while the patient table or couch moves through the gantry aperture. |
|
What happens when an x-ray tube reaches it maximum heat value
|
it will not operate until it cools down.
|
|
Do CT x-ray tubes need a high heat capacity
|
yes, CT x-ray tubes must possess a high heat capacity which is the amount
of heat that a tube can store without operational damage to the tube. |
|
What causes the heat to build up in the x-ray tube
|
The x-ray
tube must be designed to absorb high heat levels generated from the high speed rotation of the anode and the bombardment of electrons upon the anode surface. |
|
What is the approximate MHU of modern CT units
|
Modern CT systems
utilize x-ray tubes that have a heat capacity of approximately 3.5 to 5 million heat units(MHU). |
|
How do the CT scanners dissapate heat
|
oil and air cooling system
|
|
How does the CT x-ray tube anode dissepated heat
|
A CT x-ray
tube anode has a large diameter with a graphite backing. The large diameter backed with graphite allows the anode to absorb and dissipate large amounts of heat. |
|
What determines the focal spot size of the x-ray tube
|
The focal spot size of an x-ray tube is determined by the size of the filament
and cathode which is determined by the manufacturer. |
|
Do most x-ray tubes have more than 1 focal spot
|
yes
|
|
How is 2D data obtained in CT radiography
|
The fundamental principle behind computed tomography is to acquire multiple views of an object over a range of angular orientations. By this means, additional dimensional data are obtained in comparison to conventional X-radiography, in which there is only one view.
|
|
Describe how the CT anode produces a focal spot
|
The focal spot track is the surface area of the anode that is bombarded by the electron beam. In stationary anode tubes, it is a small area with dimensions of a few millimeters. In the rotating anode tube, the focal spot track is much larger because of the movement of the anode with respect to the electron beam. The figure below shows a small section of a rotating anode.
|
|
What is a benefit of a small focal spot
|
From the standpoint of producing x-ray images with minimum blur, a small focal spot is desired.
|
|
What is a disadvantage of a small focal spot
|
a small focal spot tends to concentrate heat and give the focal spot track a lower heat capacity
|
|
What happens to Kv and Ma settings if a small focal hot spot is used
|
you use low power
|
|
What is bigger the focal spot being bombarded by an electron beam or the projected focal spot spot
|
Notice that the actual focal spot, the area bombarded by the electron beam, is always larger than the projected, or effective, focal spot.
|
|
What is the size of the projected focal spot size propotional to
|
For a given anode angle, the width of the focal spot track is directly proportional to the size of the projected spot
|
|
What determines the projected focal spot angle
|
he actual relationship between focal spot width (and heat capacity) and the size of the projected focal spot is determined by the anode angle
|
|
What is the range of the anode angle
|
Anode angles generally range from about 7° to 20
|
|
Are the track width and heat capacity are inversely related to anode angle
|
yes
|
|
What is a limitation to a small anode angle
|
Although anodes with small angles give maximum heat capacity, they have specific limitations with respect to the area that can be covered by the x-ray beam
|
|
What is the heel effect
|
X-ray intensity usually drops off significantly toward the anode end
|
|
http://www.sprawls.org/ppmi2/XRAYHEAT/#Focal Spot Size
|
http://www.sprawls.org/ppmi2/XRAYHEAT/#Focal Spot Size
|
|
What is a limitation to a small anode angle
|
Although anodes with small angles give maximum heat capacity, they have specific limitations with respect to the area that can be covered by the x-ray beam
|
|
What is the heel effect
|
X-ray intensity usually drops off significantly toward the anode end
|
|
http://www.sprawls.org/ppmi2/XRAYHEAT/#Focal Spot Size
|
http://www.sprawls.org/ppmi2/XRAYHEAT/#Focal Spot Size
|
|
Is the heel effect more pronounced in tubes with small angles
|
In tubes with small angles, this is more pronounced and limits the size of the useful beam.
|
|
What does the x-ray tube look like
|
|
|
Does the speed of rotation of the anode effect the focal spot trak length of use
|
Yes, The effective length of the focal spot track is proportional to the speed of rotation for a given exposure time.
|
|
Does the anode spin like a rotor
|
yes
|
|
How does high speed rotation increase the focal trak length of use
|
High-speed rotation simply spreads the heat over a longer track, especially in short exposure times
|
|
Does the focal spot size, anode rotation speed, KV waveform and anode angle affect the heat capacity of the focal spot tract
|
yes
|
|
What is a major limiting factor of a single exposure
|
The maximum heat capacity of the focal spot area, or track, is the major limiting factor with single exposures
|
|
What will happen to the anode surface if the heat capacity is exceeded
|
If the quantity of heat delivered during an individual exposure exceeds the track capacity, the anode surface can melt, as shown below.
|
|
How does a single phase power deliver energy to the anode
|
|
|
How does a constant potential machine deliver heat
|
|
|
What is a problem with single phase power
|
Because of the pulsating nature of single-phase power, some points on the anode surface are raised to higher temperatures than others and exceed the heat capacity
|
|
Does constant potential operation increases the effective focal spot track heat capacity
|
yes
|
|
What is true of radiation dose and thin slices
|
All settings be the same thinner slices on a 64 slice machine will result in less radtion (0.625 vs 5mm), but the signal to noise decreases. Therefore to obtain high quality images with thinner slices there must be increase in Kvp or mA to compenstate and there fore the radiation does go up.
|
|
Why does a smaller phantom receive larger amount of radiation
|
|