Automatic contrast injector : also called
|
* Power injector
* Mechanical injector * Contrast medium injection devices |
|
Power injector - why use one ? (8) |
* Faster injection times = better contrast enhancement
* Large bolus injections =single large quantity * Pre-set amount * Consistent injection * Easy reproduction of injection ( settings) * Programmable * Can be used for many indication |
|
Configuration |
* Pedestal * Table bracket * Wall mount * Ceiling mount |
|
Injector Consists of: (5) |
* Control panel
* Syringes * Warming device * Pressure mechanism * Xray and injection control |
|
Control panel |
* Displays injection parameters
|
|
control panel parameters include (5) |
* Rise
* Total volume * Pressure * Delay |
|
features of the Syringe |
* Sterile
* 40ml-260ml * Removable * Steel, polycarbonate * Can be disposable |
|
features of the Heating device (5) |
* Placed over syringe
* Set to body temp - 37 degrees * Reduces viscosity of contrast * Maintains temperature of preheated contrast medium |
|
High Pressure mechanism - |
-electromechanical -options for numbers of turrets = * Single syringe ( head) assembly * Double syringe ( head) assembly * Detachable injector head |
|
1. Injection control
|
* Injector hand switch * Injector connected to x-ray generator * Can be controlled at injector or from near operator console |
|
Injection parameters |
1. Flow rate 2. totally volume 3. rise ( linear rise) 4. injection pressure 5. delay |
|
factors that affect flow rate |
* Procedure
* Vessel * Pt type * Nature of disease * Viscosity of contrast medium = Altered by heat or iodine content * Catheter = Length, Diameter, # of holes * Injection pressure ( PSI) |
|
flow rate |
* Rate contrast medium injected * Measure in cubic cm per second or mm per sec ( cc/sec or mm/sec) |
|
total volume |
* Amount of contrast to be delivered
* = flow rate x injection duration ( V = cc/sec X sec or ml/sec X sec) * Measured in cubic cm |
|
rise ( linear rise) |
* Time it takes to reach the desired flow rate * Measured in seconds |
|
injection pressure |
* Force needed for a specific dose of contrast medium * Measured in pounds per square inch ( psi) Range of 100 -200 psi * Size of vascular structure, - Smaller vessel, lower injection pressure * Type of catheter |
|
delay |
* Can be either injection or x ray delay * Allows for image acquisition to begin before injection of contrast * Useful in DSA where mask image required * Injection of contrast begins before the acquisition of images * Useful in lower extremity arteriography via catheter in abdominal aorta |
|
safety features (8) ( for details look in notes) |
* Function monitoring
* volume limiting device * Pressure limiting device * Acceleration regulation * Rate rise * Extravasation/ interstitial injection detection * Air embolism protection * Manual or automatic safety stop |
|
set up and operation |
in notes |