74. WHICH OF THE FOLLOWING CHARTS MAY BE CONSULTED TO DETERMINE THE SAFETY OF A SINGLE X-RAY EXPOSURE
|
TUBE RATING CHART
|
|
75. WHICH OF THE FOLLOWING CHARTS MAY BE CONSULTED TO DETERMINE THE SAFETY OF A SERIES OF X-RAY EXPOSURE
|
ANODE COOLING CURVE
|
|
76. THE PORTION OF THE IMAGE-INTENSIFIER TUBE THAT CONVERTS ELECTRON ENERGY TO VISIBLE LIGHT IS THE
|
OUTPUT PHOSPHOR
|
|
77. THE PORTION OF THE IMAGE-INTENSIFIER TUBE THAT CONVERTS VISIBLE LIGHT TO AN ELECTRONIC IMAGE IS THE
|
PHOTOCATHODE
|
|
78. THE INPUT PHOSPHOR OF THE IMAGE-INTENSIFIER TUBE CONVERTS
|
X-RAY ENERGY TO VISIBLE LIGHT
|
|
79. TOTAL BRIGHTNESS GAIN ACHIEVED USING AN IMAGE INTENSIFIER EQUALS
|
FLUX GAIN TIMES MINIFICATION GAIN
|
|
80. SINGLE-PHASE, FULL-WAVE RECTIFICATION PRODUCES
|
PULSATING DIRECT CURRENT WITH 120 PULSES PER SECOND AND 100% RIPPLE
|
|
81. THREE-PHASE, SIX-PULSE FULL-WAVE RECTIFICATION PRODUCES
|
DIRECT CURRENT WITH 13% RIPPLE
|
|
82. THREE-PHASE, TWELVE-PULSE FULL WAVE RECTIFICATION PRODUCES
|
DIRECT CURRENT WITH 4% RIPPLE
|
|
83. THE INCREASE IN AVERAGE PHOTON ENERGY WHEN USING THREE-PHASE, SIX-PULSE EQUIPMENT COMPARED WITH SINGLE-PHASE EQUIPMENT IS
|
35%
|
|
84. THE INCREASE IN AVERAGE PHOTON ENERGY WHEN USING THREE-PHASE, 12-PULSE EQUIPMENT COMPARED WITH SINGLE-PHASE EQUIPMENT IS
|
41%
|
|
85. PROGRAMS THAT DEAL WITH THE SAFE AND RELIABLE OPERATION OF EQUIPMENT AND THOSE THAT ADDRESS ALL ASPECTS OF THE DELIVERY OF RADIOLOGY SERVICES ARE CALLED, RESPECTIVELY
|
QUALITY CONTROL AND QUALITY ASSURANCE
|
|
86. EXAMPLES OF DEDICATED X-RAY EQUIPMENT INCLUDE
|
MAMMOGRAPHY UNITS, TOMOGRAPHY UNITS, AND MOBILE X-RAY MACHINES
|
|
87. THE COLLIMATOR MUST BE ACCURATE TO A LEVEL OF
|
2% OF SID
|
|
88. kVp MUST BE ACCURATE TO WITHING
|
4
|
|
89. EXPOSURE LINEARITY MUST BE ACCURATE TO WITHIN
|
10%
|
|
90. EXPOSURE REPRODUCIBILITY MUST BE ACCURATE TO WITHIN
|
5%
|
|
91. WHEN A SPINNING TOP TEST IS PERFORMED ON SINGLE-PHASE EQUIPMENT, A RADIOGRAPH EXHIBITING FOUR DOTS WOULD INDICATE
|
AN ACCURATE TIMER, IF SET ON 1/30 SECOND
|
|
92. WHEN A SPINNING TOP TEST IS PERFORMED ON THREE-PHASE EQUIPMENT, A TIMER SETTING OF 1/60 SECOND SHOULD INDICATE THE FOLLOWING ON THE RESULTANT RADIOGRAPH
|
A 6-DEGREE ARC
|
|
93. THE TEST THAT MEASURES THE ACCURACY OF ADJACENT mA STATIONS IS
|
EXPOSURE LINEARITY
|
|
94. THE TEST THAT MEASURES THE ACCURACY OF SUCCESSIVE EXPOSURE IS
|
EXPOSURE REPRODUCIBILITY
|
|
95. EFFECTIVE FOCAL-SPOT SIZE MAY BE MEASURED USING THE FOLLOWING TOOL(S)
|
SLIT CAMERA, STAR TEST PATTERN, AND PINHOLE CAMERA
|
|
96. RESOLUTION OF THE TELEVISION SYSTEM MAY BE MEASURED USING THE FOLLOWING TOOL(S)
|
RESOLUTION TEST PATTERN
|
|
97. AUTOMATIC EXPOSURE CONTROLS MAY BE TESTED USING
|
PHANTOMS
|
|
98. THE AMOUNT OF mA USED FOR FLUOROSCOPY IS
|
3-5
|
|
99. MARKS ON THE FOCAL TRACK OF THE ANODE RESULTING FROM BOMBARDMENT OF ELECTRONS ARE CALLED
|
PITTING
|
|
100. EFFECTIVE QUALITY CONTROL AND QUALITY ASSURANCE PROGRAMS ARE REQUIRED FOR ACCREDITATION BY
|
THE JOINT COMMISSION
|