• Shuffle
    Toggle On
    Toggle Off
  • Alphabetize
    Toggle On
    Toggle Off
  • Front First
    Toggle On
    Toggle Off
  • Both Sides
    Toggle On
    Toggle Off
  • Read
    Toggle On
    Toggle Off
Reading...
Front

Card Range To Study

through

image

Play button

image

Play button

image

Progress

1/24

Click to flip

Use LEFT and RIGHT arrow keys to navigate between flashcards;

Use UP and DOWN arrow keys to flip the card;

H to show hint;

A reads text to speech;

24 Cards in this Set

  • Front
  • Back

Purpose of hemodynamic monitoring

To evaluate the ability of the cardiovascular system to deliver blood flow and oxygenation to the tissues

How many lumen is the pulmonary artery catheter

Four lumen catheter

Before transporting always ask for

A current wedge

Wedge requirements

500cc IV bag of NS with a 300 mmHg pressure bag



* bag must be pressurized to 300 mmHg to overcome the resistance of the flush device in the transducer and to deliver 3mL per hour through the catheter



*Zeroing the transducer tells the monitoring system that the atmospheric pressure is "zero"



*low hemodynamic monitoring values may be caused by the transducer being above the phlebostatic axis, or by air bubbles or clots in the catheter

Low monitoring values may be from

May be from the transducer being above the phlebostatic axis, or by air bubbles or clots in the catheter

Leveling the transducer

Stopcock used to zero the system must be at the phlebostatic axis for all pressure readings


If transducer is lower it will cause inaccurately high readings


I'd transducer is higher it will cause inaccurately low readings



(Tape to their side to stay level)

Transducer rules

Most always be level with the right atrium


Iv pole


Phlebostatic axis... 4th intercostal space

Hemodynamic pressures

As cath approaches the right atrium it reflects the CVP and RA pressure


•Balloon is inflated as is approaches the RA to float the catheter through the chambers of the heart


•Aa balloon advances the waveform reflects the movement (or turbulance) of the catheter tip


•difference b/w the RV pressure and the PA pressure is reflected in the diastolic pressure

CVP and RAP

CVP = pressure in the great veins


RAP = pressure in the R atria. Measured through the proximal port of the PA catheter


CVP and RAP = reflect right ventricular end diastole pressure or preload. Normal values 2 to 8 mmHg

RVP

Right ventricular pressure (pressure in the right ventricle)


Seen during insertion of the PA catheter


Normal :


Systolic: 15 to 30


Diastolic: 0 to 8

PAP

Pulmonary artery pressure


Reflects both right and left heart pressures


Pressure in the pulmonary artery with the balloon deflated


Normal:


Systolic: 15 to 30


Diastolic: 5 to 15

Causes of increased/decreased PAP

Increased: hypervolemia, pulmonary HTN, PPV, cardiac tamponade, left ventricular failure



Decreased: hypovolemia, vasodilation

Wedge

Pulmonary artery wedge pressure


(PAWP/PCWP/Wedge)


Indirectly reflects left atrial pressure and left ventricular end diastolic pressure (LVEDP)

Obtaining wedge pressure

Wedging of the catheter is only done to obtain the wedge pressure.


Balloon is inflated with 1.5cc air and allowed to sail (wedge) into the distal branch of pulmonary artery where it's too narrow for balloon to pass.


During occlusion the catheter sensing tips "see through" the pulmonary circulation (no valves) into the left atrium giving an indirect reflection of the left atrial pressure (LAP)


Inflate balloon no longer than 15 to 30 seconds


Use only air to inflate


Never flush catheter in the wedge position


Re-zero and recalibrate the transducer system after each wedge reading


Normal values 4 to 12

CO

Cardiac output


Amount of blood ejected by the ventricle each minute


Evaluates cardiac function


Normal is 4 to 8 L/min


Measured by thermodilution



HR x SV = CO

MAP normal range

Mean arterial pressure


(2DBP + SBP)÷3


70 to 100 mmHg

CVP normal range

Central venous pressure


2 to 8 mmHg

PAD normal range

Pulmonary artery diastolic pressure


5 to 15 mmHg

PAP normal range

Pulmonary artery pressure


15 to 30 mmHg

Wedge normal range

PCWP / PAWP / Wedge pressure


4 to 12 mmHg

CO normal range

Cardiac output


4 to 8 L/min

CI normal range

Cardiac index


Cardiac output ÷ BSA


2.5 to 4 L/min

RVP normal ranges

Right ventricular pressure


Systolic 15 to 30 mmHg


Diastolic 0 to 8 mmHg

If dampened waveform seen

Cath might have spontaneously migrated forward. Encourage pt to cough, reposition and if persistent wedge is suspected and unable to be reduced pull catheter back until the issue has resolved into a PA waveform.



When balloon is not inflated be prepared to pull catheter back.