WINDOW WIDTH
|
ABSORPTION MEASUREMENT RANGE IN HU. MAX. # SHADES OF GRAY THAT CAN BE DISPLAYED ON MONITOR.
|
|
WINDOW LEVEL
|
CENTER OF RANGE OF CT #'S AND CAN BE POSITIONED ANYWHERE ON WW.
|
|
CT # FOR BONE
|
+1000
|
|
CT # FOR AIR
|
-1000
|
|
CT # FOR WATER
|
0
|
|
5 THINGS THAT AFFECT IMAGE QUALITY
|
SPATIAL RESOL., CONTRAST RESOL.,NOISE, RAD DOSE, IMAGE ARTIFACTS
|
|
SPATIAL RESOLUTION
|
DEGREE OF BLURRING ON IMAGE. MEASURE OF ABILITY TO DISCRIMINATE OBJECTS OF VARYING DENSITY IN A SMALL DISTANCE APART AGAINST UNIFORM BACKGROUND.
|
|
FOCAL SPOT SIZE VS SPATIAL RESOL.
|
FSS INCREASES, SR DECREASES. DETAILS IN OBJECT ARE SPREAD OUT OVER SEVERAL DETECTORS.
|
|
APERTURE SIZE
|
WIDTH OF APERTURE AT DETECTOR. HIGH SP WITH SMALL APERTURES
|
|
PARTIAL VOL. EFFECT
|
GETTING AN INCORRECT CT # (4MM OBJECT W/ SLICE THICKNESS OF 10MM SPREAD THE 4MM OVER ENTIRE SLICE THICKNESS)
|
|
# OF PROJECTIONS VS. SP
|
# OF PROJ INCREASES, SR INCREASES (MORE DATA FOR IMAGE RECON)
|
|
IMAGE RECONSTRUCTION INVOLVES 2 MATH PROCEDURES
|
CONVOLUTION AND BACK PROJECTION
|
|
CONVOLUTION (KERNEL)
|
AFFECTS APPEARANCE OF IMAGE STRUCTURES. HIGH-PASS FILTER.
|
|
PIXEL SIZE =
|
FOV/MATRIX SIZE
|
|
LOW CONTRAST RESOLUTION
|
ABILITY OF CT UNIT TO DEMONSTRATE SMALL CHANGES IN TISSUE CONTRAST.
|
|
FACTORS AFFECTING LOW CONTRAST RESOLUTION
|
PHOTON FLUX, SLICE THICKNESS, PT SIZE, DETECTOR SENSITIVITY, RECON ALGORITHM, IMAGE DISPLAY, IMAGE RECORDING, QUANTUM NOISE
|
|
SOURCES OF NOISE
|
QUANTUM NOISE(# OF DETECTED PHOTONS), MATRIX SIZE, SLICE THICKNESS, ALGORITHM, ELECTRONIC NOISE (DETECTOR ELECTRONICS), SCATTERED RADIATION, OBJECT SIZE
|
|
TYPES OF IMAGE ARTIFACTS
|
STREAKS, SHADING, RINGS AND BANDS
|
|
RECORD PT DOSE
|
IONIZATION CHAMBER (CTDI)
|